User Contributed MET/CAL PROCEDURE ============================================================================= INSTRUMENT: Keithley 169: (1 Year) CAL VER DATE: 30-Apr-93 AUTHOR: User Contributed REVISION: 1.00 ADJUSTMENT THRESHOLD: 70% NUMBER OF TESTS: 31 NUMBER OF LINES: 138 CONFIGURATION: Fluke 5700A ============================================================================= STEP FSC RANGE NOMINAL TOLERANCE MOD1 MOD2 3 4 CON 1.001 ASK- R F W 1.002 ASK+ B 1.003 MMFC C1 # # 1.004 HEAD * * * * * * * * * CONNECTION DATA * * * * * * * * * 1.005 5700 * S 1.006 DISP ! Connect the 5700A and the 169 as follows. 1.006 DISP ![27][91]1m 5700A TO 169[27][91]m 1.006 DISP ! OUTPUT VêA ------------> INPUT HI 1.006 DISP ! OUTPUT LO ------------> INPUT LO 1.006 DISP 1.006 DISP Turn the 169 on and check for low battery indication. 1.006 DISP If the low battery indicator (BAT) is on, remove and 1.006 DISP replace the batteries with fresh ones. # # 1.007 HEAD * * * * * * * * * * * WARNING * * * * * * * * * * * 1.008 DISP WARNING !! Some procedures require the use of high 1.008 DISP voltage. Take care to prevent contact with live 1.008 DISP circuits which could cause electrical shock resulting 1.008 DISP in injury or death. # # 1.009 HEAD { INITIAL FUNCTION CHECK } 1.010 DISP Power on, all other switches out. (-) may flash. 1.011 EVAL Does the display read 000 ? 2.001 DISP Select the ACV Function and 200mV range. 2.002 5700 0Z S RW 2.003 EVAL Does the display read 00.0 ACmV 3.001 DISP Select the DCV Function and 200mV range. 3.002 5700 0Z S RW 3.003 EVAL Does the display read 00.0 DCmV 4.001 DISP Press A & ê at the same time 4.001 DISP then select the 2Kê range. 4.001 DISP Display reads 2 mA current shunt on 2Kê 4.001 DISP range and checks the current fuse 4.001 DISP Note: Display reading may vary in last position! 4.002 EVAL Does the display read .100 mAKê 5.001 DISP Press 200ê & 2m at the same time 5.001 DISP Display reads 200mA current shunt on 200ê 5.001 DISP range . ( About 100ês ) 5.001 DISP Note: Display reading may vary in last position! 5.002 EVAL Does the display read 1.00.0 uAê 6.001 DISP Press 200ê & 20mA at the same time 6.001 DISP Display reads 200mA current shunt on 200ê 6.001 DISP range. 20mA shunt 200ê range. ( 10ê ) 6.001 DISP Note: Display reading may vary in last position! 6.002 EVAL Does the display read 1.0.0 uAê 7.001 5700 0Z S RW 7.002 DISP Select the ês Function and 20M range 7.003 EVAL Does the display read 0.00 Mê 8.001 5700 0Z S RW 8.002 DISP Select the 2000K range 8.003 EVAL Does the display read 000 Kê 9.001 EVAL Does the display read 1 Kê # # 10.001 HEAD { DC VOLTAGE PERFORMANCE CHECK } 10.002 ASK+ R 10.003 DISP Select the DC Volts function. 10.004 5700 200 100mV 0.4U 2W #! Test Tol 0.0004, Sys Tol 1.45e-006, TUR 275.862 (>= 4.00). 11.001 5700 2 1V 0.004U 2W #! Test Tol 0.004, Sys Tol 7.2e-006, TUR 555.556 (>= 4.00). 12.001 5700 20 10V 0.04U 2W #! Test Tol 0.04, Sys Tol 5.4e-005, TUR 740.741 (>= 4.00). 13.001 5700 200 100V 0.4U 2W #! Test Tol 0.4, Sys Tol 0.0007, TUR 571.429 (>= 4.00). 14.001 5700 1000 1000V 4U 2W #! Test Tol 4, Sys Tol 0.0086, TUR 465.116 (>= 4.00). # # 15.001 HEAD { AC VOLTAGE PERFORMANCE CHECK } 15.002 DISP Select the AC Volts function. 15.003 5700 200 100mV 1.3U 1kH 2W #! Test Tol 0.0013, Sys Tol 2e-005, TUR 65.000 (>= 4.00). 16.001 5700 2 1V 0.013U 1kH 2W #! Test Tol 0.013, Sys Tol 8.2e-005, TUR 158.537 (>= 4.00). 17.001 5700 20 10V 0.13U 1kH 2W #! Test Tol 0.13, Sys Tol 0.00082, TUR 158.537 (>= 4.00). 18.001 5700 200 100V 1.3U 1kH 2W #! Test Tol 1.3, Sys Tol 0.009, TUR 144.444 (>= 4.00). 19.001 5700 1000 1000V 13U 1kH 2W #! Test Tol 13, Sys Tol 0.084, TUR 154.762 (>= 4.00). # # 20.001 HEAD { RESISTANCE PERFORMANCE CHECK } 20.002 DISP Select the ê Function. 20.003 5700 200 190Z 0.9U RW #! Test Tol 0.9, Sys Tol 0.00523, TUR 172.084 (>= 4.00). 21.001 5700 2 1.9kZ 0.005U RW #! Test Tol 5, Sys Tol 0.0328, TUR 152.439 (>= 4.00). 22.001 5700 20 19kZ 0.05U RW #! Test Tol 50, Sys Tol 0.309, TUR 161.812 (>= 4.00). 23.001 5700 200 190kZ 0.5U 2W #! Test Tol 500, Sys Tol 2.47, TUR 202.429 (>= 4.00). 24.001 5700 2000 1900kZ 5U 2W #! Test Tol 5000, Sys Tol 36.1, TUR 138.504 (>= 4.00). 25.001 5700 20 10MZ 0.21U 2W #! Test Tol 2.1e+005, Sys Tol 370, TUR 567.568 (>= 4.00). # # 26.001 HEAD { DC CURRENT PERFORMANCE CHECK } 26.002 DISP Select the DC Amp Function. 26.003 5700 200 100uA 0.9U 2W #! Test Tol 9e-007, Sys Tol 1.5e-008, TUR 60.000 (>= 4.00). 27.001 5700 2 1mA 0.009U 2W #! Test Tol 9e-006, Sys Tol 6e-008, TUR 150.000 (>= 4.00). 28.001 5700 20 10mA 0.09U 2W #! Test Tol 9e-005, Sys Tol 6e-007, TUR 150.000 (>= 4.00). 29.001 5700 200 100mA 0.9U 2W #! Test Tol 0.0009, Sys Tol 7e-006, TUR 128.571 (>= 4.00). 30.001 5700 2000 1000mA 9U 2W #! Test Tol 0.009, Sys Tol 0.00011, TUR 81.818 (>= 4.00). # # 31.001 HEAD * * * * 169 VERIFICATION COMPLETED * * * * 31.002 5700 * S 31.003 DISP Please turn the 169 Digital Multimeter. 31.004 END #! T.U.R.s less than 4.00: 0 #! T.U.R.s estimated using RANGE value: 0 #! T.U.R.s not calculated (ASK- U): 0 #! T.U.R.s not computable at compile time: 0 #! FOR JUSTIFICATION REFER TO COMMENTS FOLLOWING EACH TEST IN THIS LISTING.